Chapter 4: Running Samples 115
5 Check the fluidics levels in the Cytometer window. Replenish fluids or
empty the waste, if needed.
To display the Cytometer window, click the Cytometer button ( ) on the
Workspace toolbar. Check the fluid levels at the bottom of the window
(Figure 4-2).
Figure 4-2 Fluid levels in Cytometer window
To serv ice the f luid containers, se e Setting Up the Fluidics Car t on pag e 122
and Emptying the Waste Container on page 127.
Performing Fluidics Startup
This section describes the fluidics startup procedure. The status of the fluidics
system is displayed in the bottom right corner of the main window.
1 From the BD FACSDiva Cytometer menu, select Fluidics Startup. The
following window appears.
ethanol
bleachDIwaste
sheath
Note: Roll the mouse pointer over
fluid indicators to show fluid type.