66 BD FACSAria II User’s Guide
Table 2-2 Side Stream window controls
Control Description
Voltage button Turns the plate voltage on and off.
Test Sort button Generates test side streams based on test sort pulses.
Optical Filter button Controls the position of the optical filter in front of the lower
(Accudrop) camera.
Attenuation button Decreases the amplitude of the drop drive. At lower pressures,
you may need to turn on attenuation to dampen the amplitude.
Waste Drawer button Opens or closes the aspirator drawer depending on its current
state. The default state is closed. For more information, see
Aspirator Drawer on page 30.
Drop Delay field Sets the amount of time between when an event is measured
and the breakoff point, from 10–140 drops. The drop delay
value determines which drop will be deflected.
The drop delay value is set experimentally using BD FACS
Accudrop technology.
Auto Delay Opens the Auto Drop delay dialog.
Voltage sliders (far
left, left, right, far
right)
Set the percentage of charge to be applied to the corresponding
stream (as a percentage of maximum).
Voltage Center slider Adjusts the relative voltage between the left and right plates
which moves the streams slightly to the left or right as a whole.
In general, the slider rarely needs adjusting.
on
off
on off
in out
off on
closed open