190 BD FACSAria II User’s Guide
Fluidics Shutdown
The Fluidics Shutdown command can be used to perform an extensive cleaning if
the system is used to process many different sample types, or the system has
problems with contamination. This procedure removes sheath fluid from the
lines and fills them with 70% ethanol, and cleans the flow cell.
Preparing for Shutdown
1 Unload the sample tube, if one is loaded.
2 Turn off the stream.
3 Check the waste container and empty it if needed.
See Refilling the Sheath Tank on page 123.
4 Check the ethanol shutdown tank and refill if it needed.
See Refilling the Ethanol Shutdown Tank on page 125.
Running Fluidics Shutdown
1 Select Cytometer > Fluidics Shutdown. The following window opens.
2 Remove the nozzle from the flow cell assembly and click Done.
3 Insert a closed-loop nozzle into the flow cell assembly and click Done. See
Figure 6-1.