98 BD FACSAria II User’s Guide
•LoadLifts a tube into the sample injection chamber, starts sample
agitation (if agitation is turned on), and starts acquisition of the sample.
When a tube is already loaded, the button changes to Unload. Clicking
Unload stops acquisition of the sample, turns off the agitator, and lowers
the tube from the sample injection chamber.
NOTE The Load button is enabled only when the workstation is connected
to the cytometer, an experiment is open, the stream is turned on, and the
current tube pointer is set to a tube.
Stop Acquiring—Stops sample acquisition without unloading the sample
tube.
The Acquire Data button functions only when a tube is loaded.
Flow Rate—Controls sample flow rate, from 1.0–11.0 (10–80 µL/min).
Do one of the following to change the flow rate:
- Select the value in the field and enter a new value.
- Click the pointer in the slider bar and drag it to a new value.
- Use the mouse to click the up and down arrows or press the arrow keys
on your keyboard to increase or decrease the values in small
increments.
Sorting Controls
All sorting on the BD FACSAria II cytometer is controlled by BD FACSDiva
software. Sorting controls are shown in Figure 3-8 on page 99. See Table 3-1 on
page 99 for explanations of each area.