198 BD FACSAria II User’s Guide
To run t he Prepare fo r Aseptic Sort command:
1 Select Cytometer > Cleaning Modes > Prepare for Aseptic Sort.
Follow the instructions on screen as you perform the procedure.
2 Install the closed-loop nozzle in the flow cell and click Done.
Verify that there is an O-ring in the closed-loop nozzle before installing it.
3 Remove the fluid line from the DI water port, connect the fluid line from
the bleach tank to the DI water port, then click Done. See Figure 6-4.
Do not disconnect the sensors from either tank.
Figure 6-4 Moving bleach fluid line
DI water line
disconnected
bleach line connected to
DI water port