Aria 8.0 Betriebsanweisung Seite 293

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 344
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 292
Appendix C: BD Aerosol Management Option 293
Figure C-9 Seating the new filter
NOTE For optimal evacuation of aerosols, the filter must be completely
seated in the evacuator filter well.
7 Lift the spring-loaded filter hold-down and place it on top of the filter.
8 Press and hold the Filter Life Reset button on the membrane panel for
5–10 seconds.
The green and amber LED lights will turn off, and then on. Hold down the
button until the 100% indicator light is lit. This resets the 180-hour filter
life clock.
9 Connect one end of the replacement tubing to the ULPA filter, and the
other end to the tubing manifold (Figure C-10 on page 294).
Reset button
Seitenansicht 292
1 2 ... 288 289 290 291 292 293 294 295 296 297 298 ... 343 344

Kommentare zu diesen Handbüchern

Keine Kommentare