318 BD FACSAria II User’s Guide
Figure E-3 Creating an experiment from the QC template
4 Add the current month and year to the experiment name, rename the
specimen with the type of bead you are running (such as Rainbow beads),
and rename the tube with today’s date.
Adjusting Area Scaling for the Blue Laser
1 Open the QC experiment by double-clicking the experiment icon.
Verify that the Area Scaling worksheet is displayed in the Worksheet
window.
2 Click to set the current tube pointer next to the tube with today’s date in
the Browser.
3 Install the tube of beads onto the loading port.
4 Click Load in the Acquisition Dashboard.
current tube
pointer