Chapter 6: Shutdown and Maintenance 195
Sample Line Backflush
After a sample tube is unloaded, the sample line tubing within the sample
injection chamber is automatically flushed inside and out with sheath fluid to
eliminate potential sample carryover. Use the Sample Line Backflush command to
perform additional backflushing of the inside of the sample line after a tube is
unloaded. Perform the Sample Line Backflush when you observe sample
carryover or after running samples with adherent cells or dye.
NOTE Keep the stream running while performing the backflush.
1 Select Cytometer > Cleaning Modes > Sample Line Backflush.
Table 6-2 Cytometer cleaning modes
Cleaning Mode Summary Frequency
Sample Line Backflush on
page 195
Flushes the sample line with
sheath fluid
After running samples
with adherent cells or dye
Clean Flow Cell on
page 188
Cleans the sample path and
the flow cell with a cleaning
fluid
When indicated by
distorted scatter or high
CVs, or as a daily
shutdown procedure.
Prime After Tank Refill on
page 196
Primes the fluid lines for the
designated fluid(s)
When a fluidics line is
unplugged to refill a tank
Prepare for Aseptic Sort
on page 197
Decontaminates the complete
sheath path and sample path
with bleach, DI water, and
ethanol.
When needed before
aseptic sorting