212 BD FACSAria II User’s Guide
Changing the Sheath Tank Air Filter
Check the inline air filter on the sheath tank air line periodically for any signs of
debris or discoloration. Replace with a new air filter from the accessory kit every
six months, or sooner if needed.
Figure 6-16 Inline air filter for sheath tank
1 Turn off the cytometer.
2 Pull the tubing off each end of the air filter.
3 Install a new filter with the directional arrow pointing toward the sheath
tank.
Checking the Fluidics Cart Drip Tray
Check the drip tray on the fluidics cart weekly for signs of excess liquid and
empty the tray if needed.
sheath tank inline
air filter
drip tray