6 Pull the sample line up to operation height, slightly above the chamber
viewing window.
7 Place a tube onto the loading port.
8 Click Done, and the sample injection chamber raises.
The sample line should not bow or bend when a tube is loaded.
9 Adjust the sample line height if needed.
10 Tighten the sample line fitting and click Done.
The bulk injection chamber is lowered and the sample purge mode is
turned on for about 5 seconds until several drips exit the filter. A message is
displayed at the bottom of the wizard. See Figure 6-24 on page 222.
Figure 6-24 Sample line wizard message
11 Click Done and the process is complete.
Be careful of a pinching hazard as the sample injection chamber is raised.