Aria 8.0 Betriebsanweisung Seite 123

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 344
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 122
Chapter 4: Running Samples 123
NOTE Make sure the blue sheath fluid line between the sheath tank and the
cytometer does not come into contact with anything that could introduce
vibration that might affect the stability of the stream.
Refilling the Sheath Tank
The fluid level in the pressurized stainless steel sheath tank should be checked
often and refilled when low. The startup times for the breakoff stream will
increase when the tank is low.
NOTE If the cytometer is run until the sheath supply is too low, the system will
turn off the stream.
Figure 4-4 Sheath tank connectors
1 Turn off the stream.
2 Disconnect the air line.
3 Pull up on the ring of the pressure relief valve to release pressure from the
tank.
Make sure the tank is fully vented.
4 Unscrew the sheath tank cover knob and remove the cover.
sheath fluid
connector
sheath sensor
sheath tank
cover knob
pressure relief
valve
air connector
sheath filter
Seitenansicht 122
1 2 ... 118 119 120 121 122 123 124 125 126 127 128 ... 343 344

Kommentare zu diesen Handbüchern

Keine Kommentare