Appendix E: QC Using BD FACSDiva Software 315
The steps in this section show you how to set up the cytometer configuration in
preparation for performing QC.
1 Select Cytometer > View Configurations, which opens the Cytometer
Configuration window in the CS&T module. See Figure 3-5 on page 86.
NOTE When you select Cytometer > View Configurations, the cytometer
disconnects from the BD FACSDiva interface and connects to the CS&T
interface. While the CS&T module is running, BD FACSDiva is in a
holding mode and does not accept any user input. When CS&T is closed,
BD FACSDiva becomes active again.
2 Select a cytometer configuration that contains appropriate settings for the
samples you are running.
3 Once the appropriate configuration is selected, click OK in the Cytometer
Configuration window to close it.
4 Close the CS&T application by selecting File > Exit. The BD FACSDiva
interface is now active and accepting user input.
See Figure E-1 and Figure E-2 for error messages that can occur if the sort
setup or sheath pressure are different between the cytometer configuration
in CS&T and the values in BD FACSDiva.
Figure E-1 Possible error message when closing CS&T
For accurate data results, the cytometer optics must match the current
cytometer configuration. Also, the sheath pressure and sort setup must
match the settings in BD FACSDiva.