Chapter 6: Shutdown and Maintenance 227
3 Open the sort block door.
4 Place 1–2 drops of DI water or ethanol on a cotton swab.
5 Click the Breakoff window and select Raw Image.
Visible smudges are more apparent in the raw image view.
6 While viewing the image on the Breakoff window, insert the swab just
below the bottom of the flow cell.
The strobe lens and upper camera windows are located behind two circular
openings on either side of the top of the sort chamber. You will see the end
of the swab in the Breakoff window when you intercept either opening.
7 Gently wipe the upper camera window, and then the strobe lens (opposite
the window) to remove any saline.
8 Repeat with isopropyl alcohol or 70% ethanol until clean.
upper
camera
window
strobe lens
window