280 BD FACSAria II User’s Guide
Troubleshooting
Observation Possible Causes Recommended Solutions
Signals dim or noisy Wrong laser turned on
(Near UV instead of violet,
or vice-versa).
Turn laser power switch to
correct position.
Laser delay set incorrectly
(possible holdover from
setting on other laser).
Recalibrate laser delay.
No signal Laser shutter closed
(presumably because flow
cell access door is open).
Close flow cell access door.
Laser power switch set to
OFF.
Turn switch to Near UV.
Laser power switch set to
violet laser instead of Near
UV.
Turn switch to Near UV.
No signals for side
population
application
Wrong filters installed. Install filters for side
population application.