Appendix B: Near UV Laser Option 279
6 Adjust the voltages so that the red blood cells are seen in the lower left
corner and the dead cells line up on the vertical line to the far right.
7 The bulk of the rest of the cells can be centered. It should be possible to
identify a major G
0
/G
1
population with S/G
2
.
8 Collect a file of 50,000 to 100,000 of live gated events.
SP prevalence is ~0.02 to 0.05% of total.
9 If sorting, sort into tubes/plate containing small volume of DMEM
containing 10% FBS and 10 mmol HEPES. Keep sample chilled at 4
°C.
; Tip Sort rare event samples into polypropylene tubes. Avoid polystyrene
tubes.
Figure B-5 Typical plot from side population experiment