Chapter 5: Sorting 177
; Tip Click Record Data to save data for the tube. Acquisition and sorting
continue when the required number of events has been recorded.
Sorting continues until the required number of cells has been sorted.
Acquisition stops and the drawer closes when sorting is complete. If the
number of Target Events is set to Continuous, sorting continues until you
manually stop sorting by clicking the Stop Acquiring button in the
Dashboard, or the Sort button in the Sort Layout window.
Monitor the sort progress from the Sort Layout window. The number of
events sorted into each sort location appears in the corresponding field.
The sort rate and sort conflict rate are displayed in the corresponding
counter fields.
NOTE When the Sweet Spot is on, sorting pauses automatically if the
Drop 1 or Gap values are out of range. This ensures that sorting occurs
only under the proper breakoff conditions. If a more severe problem such
as a clog is detected, the stream shuts off and sorting stops. The deflection
plates shut off, the aspirator drawer closes, and the sample tube is
unloaded. To clear a clogged nozzle, see Cleaning the Nozzle on page 216.
7 (Optional) You can print the Sort Report at this time, or open the report
later and print it then. You can also export the report.