304 BD FACSAria II User’s Guide
Figure D-2 Input and output ports on water bath
5 Connect the insulated hoses from the recirculating water bath to the
cytometer control panel.
NOTE Make sure you match the flow direction from the water bath port
to the flow direction on the instrument manifold. The instrument ports are
labeled “cooling in” and “cooling out.”
Figure D-3 Connections for temperature control option
6 Fill the water bath with distilled water containing 0.1 g/L sodium
carbonate.
Sodium carbonate helps reduce corrosion. Refer to the water bath
manufacturer’s documentation for fill levels and other setup information.
7 Plug in the water bath power cord.
Cooling OutCooling In