Chapter 3: Using BD FACSDiva Software 81
Sample Agitation
Select Cytometer > Sample Agitation to specify the speed at which samples are
agitated. You can select from one of the specified values, or select Off to turn off
agitation.
Each sort setup option is optimized at a preset sheath pressure. If you
change the sheath pressure, a multitude of other values will be affected and
need updating, including the drop drive frequency, drop charge levels, laser
delay, area scaling factor, and other values. For proper cytometer operation,
change the sheath pressure by choosing an option from the Sort > Sort
Setup menu. Do not adjust the pressure using controls in the Sheath
Pressure dialog. Note that the Sheath Pressure command is disabled when
the Sweet Spot is on.