Appendix B: Near UV Laser Option 277
Switching Back to the Violet Laser
NOTE Depending on the application, the filters for the standard violet laser can
also be used with the Near UV laser.
This section provides an overview of the main steps.
1 Place the UV/Off/Violet laser switch in the Violet position.
This turns off the Near UV laser and turns on the violet laser. Allow the
violet laser to warm up for 30 minutes before recording data.
2 Set the cytometer configuration to a configuration with violet laser
parameters.
3 Install the correct optical filters the violet laser in the far-left trigon.
4 Make sure the cytometer has a valid performance check or perform QC.
Verify the configuration contains the correct parameters for use with the
violet laser. The system will not perform correctly if this is not done.
Detector Array
(Laser)
PMT
LP Mirror
BP Filter
Replaceme
nt
Part No.
Intended Dye
Trigon
(405-nm violet
laser)
A502343797Alexa Fluor
® 430
530/30 343798
B450/40343801Cascade Blue
®,
Pacific Blue
™,
Hoechst, DAPI,
Alexa Fluor
® 405