Chapter 4: Running Samples 137
5 Select the H checkbox to select Height for each parameter. See Figure 4-12
on page 137.
Figure 4-12 Inspector Window - Parameters tab
Adjusting Area Scaling
The required area scaling factor changes based on sheath pressure and particle
size. The area scaling factors should be verified for each experiment performed
on the cytometer.
1 In the Browser, right-click Global Sheet1 and select Apply Analysis
Template.
2 In the Template dialog, select the Area Scaling worksheet and click OK.
3 Create a new specimen by clicking the New Specimen button on the
Browser toolbar.
4 Expand the new specimen, then click to set the current tube pointer to
Tube_001.
5 Install the FITC-positive control tube onto the loading port and click Load
in the Acquisition Dashboard.