Appendix D: Temperature Control Option 305
NOTE Do not start up the water bath until after you have connected the
recirculating water tubing, as described in the following sections.
Setting Up the Tube Holder
This section describes how to install the temperature control tube holder on the
instrument and how to attach the recirculating water tubing to the tube holder.
1 Place collection tubes in the temperature control tube holder.
Tube holders are available for 15- mL two-way sorting a nd 12 x 75-mm
four-way sorting. For compatible tubes, refer to Appendix A.
2 Attach the recirculating water tubing to the tube holder.
Close the sort block door and open the sort collection chamber door, if
needed. (The sort block door must be closed in order to open the collection
chamber door.)
Note that the tubing ports are labeled In and Out. Attach the input tubing
to the port on the left side of the collection tube holder, and the output
tubing to the port on the right side, as shown in Figure D-4 on page 306.
To a ttach the tub ing, pus h it into the port until the tubing sna ps into place.
Any instrument surface that comes in contact with biological specimens can
transmit potentially fatal disease. Use universal precautions when handling
sorting hardware. Wear suitable protective clothing and gloves.