194 BD FACSAria II User’s Guide
Scheduled Maintenance
For optimal cytometer functioning, perform the following procedures according
to the recommended maintenance schedule.
Internal Cleaning
BD FACSDiva software includes four pre-programmed cleaning modes that can
be used alone, or in combination to provide the required level of cleaning. The
following sections describe the different cleaning modes. See Table 6-2 on
page 195 for an overview of each mode.
Table 6-1 Scheduled Maintenance
Procedure Recommended Frequency
Internal Cleaning on this page See Table 6-2 on page 195
Purging the Fluid Filters on page 200 Weekly
Purging the Sheath Filter on page 200 Weekly
Changing the disposable waste cap (see
Emptying the Waste Container on
page 127)
Monthly
Flow Cell Preventive Cleaning on
page 201 (with 100% Conrad70)
Monthly
Changing Fluid Filters on page 202 Every 6 months
Changing the Sheath Filter on page 203 Every 3 months
Changing the Sample Lines on page 204 Every 4–6 months, or when required
Changing the Air Filters on page 211 Every 6–12 months, depending on
cytometer use and the quality of the air
Changing the Sheath Tank Air Filter on
page 212
Every 6 months
Checking the Fluidics Cart Drip Tray on
page 212
Weekly