288 BD FACSAria II User’s Guide
Opening the Sort Collection Chamber Door
1 Use one of the following options to stop sample flow into the sort
collection chamber.
•Click the Move Drawer control to close the aspirator drawer and sort
sample into waste.
•Click the Acquire button to stop sample acquisition without unloading
the sample tube.
•Click Unload to stop acquisition and unload the sample tube.
2 Press the up arrow button repeatedly to increase the suction to 100%.
3 Wait at least 30 seconds before opening the sort chamber doors to allow
potentially hazardous aerosols to evacuate.
4 Place the system in standby by pressing the POWER button on the
membrane panel of the evacuator.
5 Open the sort collection chamber door.
; Tip Remember to decrease the suction control rate to 20% when you return to
sorting samples.
Cell sorters that use droplet generation methods, like the BD FACSAria II,
can produce aerosols around the sample stream. When acquiring
biohazardous samples, follow universal precautions at all times. Keep the
sort block door and the sort collection chamber door closed during sorting.
Follow these steps to stop sample flow and evacuate potential aerosols
before opening the sort collection chamber door.