Aria 8.0 Betriebsanweisung Seite 326

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 344
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 325
326 BD FACSAria II User’s Guide
2 Adjust the PMT voltage for each parameter to set the signal to
approximately 100 x 10
3
.
3 Set the Flow Rate to 1.0 in the Dashboard.
4 Ensure that the current tube pointer is still set to the first tube, and then
click Record Data.
Once the tube has finished recording, the tube icon changes to a tube with a
disk. Also, the tube will now have cytometer settings associated with it.
5 After data has been recorded, click Unload in the Dashboard and remove
the tube.
6 (Optional) Select File > Print.
Keep a record of the daily QC results for future reference. By comparing
results from day to day, you will be able to monitor instrument
performance.
An example of recorded data is shown in Figure E-6.
tube with data
tube without data
Seitenansicht 325
1 2 ... 321 322 323 324 325 326 327 328 329 330 331 ... 343 344

Kommentare zu diesen Handbüchern

Keine Kommentare