290 BD FACSAria II User’s Guide
•Do not disconnect the tubing from the instrument manifold outlet or the
ULPA filter unless you are changing the filter. Repeated removal and
reattachment of the tubing could loosen the connection and disrupt airflow.
•To ensure optimal airflow, keep the tubing free of kinks and away from
sharp or heavy objects. Do not crush or puncture the tubing. Ensure that
the tubing is securely attached at both ends before turning on the evacuator
power.
•Keep the sort collection chamber free of potentially obstructive debris, such
as Kimwipes® or disposable pipettes.
Replacing the ULPA Filter
Replace the filter when either of the following conditions occur:
The filter-flow gauge reads 2.4 inches of H
2
O or greater at 20% suction.
The red filter-life indicator LED is blinking.
NOTE When the red light turns on and is not blinking, you have approximately
1 hour of filter life remaining. If the red light comes on during sorting, the filter
will not stop working. Replace the filter as soon as possible when the red light
starts blinking.
To replace the ULPA filter, do the following:
1 Turn off the evacuator main power and disconnect the electrical plug.
2 Disconnect the tubing from the manifold (Figure C-6 on page 291).
The manifold is located in the connection panel on the left side of the
instrument.
To p revent potential shock, always turn off the evacua tor main power and
disconnect the electrical plug from the power source before installing or
removing any filter.