174 BD FACSAria II User’s Guide
Setting Up the Experiment
;
Tip When more than one drop is deflected in the same direction, residual charge
from the first drop degrades the quality of the side streams. Thus, when four-way
sorting or sorting into small wells where precise deflection is required, use the 4-
Way Purity mode or select a mode with a Yield Mask of zero. For more
information, see Yield Mask on page 70.
1 Create a new sort layout by clicking the New Sort Layout button ( ) on
the Browser toolbar.
By default, the 2-Tube Sort Layout appears.
2 Make appropriate entries in the Sort Layout window.
• Select the collection device from the Device menu.
• Change the sort precision mode to Purity (two tubes), 4-Way Purity
(four tubes), or Single Cell (plate or slide).
•Enter the number of target events by choosing a value from the menu
or entering a number in the field.
• Select a Save Sort Reports option: Save None, Save All, or Ask User.
See Setting Up a Sort Layout on page 105.
• Select the Save Conflicts checkbox if you are using a 2- or 4-Tube Sort
Layout and want to save conflicts.