Appendix C: BD Aerosol Management Option 289
Turning Off the Evacuator
Turn off the evacuator after you have finished running biohazardous samples.
1 Place the system in standby by pressing the POWER button on the
membrane panel of the evacuator.
2 Switch off the main power on the back of the evacuator (Figure C-4 on
page 286).
Maintenance
Use the following guidelines to ensure optimal performance of the BD Aerosol
Management Option.
•Change the ULPA filter and attached tubing when the flow gauge indicator
is >2.4 at a 20% flow setting or when the red filter-life indicator LED is
blinking. See Replacing the ULPA Filter on page 290.
Two spare filters and replacement tubing are included with the AMO. To
order additional replacement kits, contact your local BD Biosciences
representative.
•Change the air filter in the sort collection chamber door on a monthly
basis. The filter traps airborne dust that could clog the ULPA filter. Regular
replacement of the air filter will extend the life of your ULPA filter. See
Replacing the Air Filter on page 294.
•Do not touch the Filter Life Reset button during normal operation. Doing
so could shut down the evacuator and prevent the collection of aerosols.
All biological specimens and materials coming into contact with them can
transmit potentially fatal disease. Handle the ULPA filter and attached
tubing as if capable of transmitting infection. Dispose of waste using proper
precautions and in accordance with local regulations. Wear suitable
protective clothing, eyewear, and gloves.