Aria 8.0 Betriebsanweisung Seite 179

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 344
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 178
Chapter 5: Sorting 179
Setting Up for Sorting Into a Plate or Slide
The following sections describe how to set up for sorting into a plate or slide. For
general guidelines, see Setting Up for Sorting on page 162.
Installing the Sorting Hardware
1 Install the splash shield below the aspirator drawer.
a Close the sort block door and open the sort collection chamber door, if
needed.
The sort block door must be closed in order to open the collection
chamber door.
b Remove the tube holder, if one is installed.
c Slide the splash shield into the slotted fittings below the sort aspirator
drawer and push it all the way in.
Any cytometer surface that comes in contact with biological specimens can
transmit potentially fatal disease. Use universal precautions when handling
sorting hardware. Wear suitable protective clothing and gloves.
splash shield
Seitenansicht 178
1 2 ... 174 175 176 177 178 179 180 181 182 183 184 ... 343 344

Kommentare zu diesen Handbüchern

Keine Kommentare