296 BD FACSAria II User’s Guide
Troubleshooting
The tips in this section are provided to help you troubleshoot issues that arise
when using the BD Aerosol Management Option. For instrument-specific
troubleshooting, refer to your instrument manual.
If additional assistance is required, contact your local BD Biosciences technical
support representative or supplier.
If any of the following are observed, assume the AMO is not evacuating
properly, and do not open the doors to the sort chamber.
Control Panel Troubleshooting
Observation Possible Cause Recommended Solution
Evacuator indicator
lights off
Evacuator power cord
unplugged
Connect the evacuator power
cord to the power source.
Evacuator power
switched off
Switch on the evacuator main
power. See Figure C-4 on
page 286.
Evacuator motor failure Contact your local BD technical
support representative.
Circuit breaker tripped Depress the circuit breaker on the
rear of the evacuator into its
original position.
Site power failure Turn off the evacuator power
switch and wait for site power to
be restored.
Evacuator indicator
lights pulsing
Erratic power source Plug the power cord into a
different outlet.