Appendix E: QC Using BD FACSDiva Software 319
The loading port rises to enclose the tube within the chamber. Once the
tube is loaded, acquisition starts automatically.
5 Set the Events to Display to 500 events.
-
6 Adjust the FSC and SSC voltages to place the particles within the P1 gate at
approximately 100 x 10
3
on scale in the FSC-H vs SSC-H dot plot.
a Click the Parameters tab in the Cytometer window.
b Select the voltage field for the parameter you need to adjust.
c To adjust the valu es you can:
Click the up and down arrows to adjust the voltages in increments
of 1.
Drag the sliders to adjust the voltages.
Press the Ctrl key while clicking the on-screen arrows to increase or
decrease in increments of 10.
Press the Ctrl key while pressing the keyboard arrows to increase or
decrease in increments of 10.